Advantages
- Flow/Immersion sensor type
- Accurate measurement
- Easy to operate & maintain
- Stable performance
Product Description
The product adopts the principle of scattering light in the direction of 90℃, and quantitatively measure the content of suspended particulate matter and the turbidity of water.
Sensor Specification
Parameter | Flow type | Immersion type |
Measurement range | 0~20;0~200;0~2000NTU | 0~100~1000~4000NTU |
Accuracy | ±2%FS | <3% |
Response time | 1s (can be adjusted) | 1s (can be adjusted) |
Operating temperature | 0~50℃ | 0~50℃ |
Operating pressure | 0.05~0.2Mpa | ≤1bar |
Shell material | PC | POM/316L |
Sensor Size | 340*280*160mm | 33*161mm |
Transmitter Specification
Turbidity Transmitter Specification |
|
Analog output | 4~20mA |
Digital output | MODBUS/RS485 |
Output relay | Optional 2-way alarm relay (250VAC, 10A) |
Power supply | 220VAC,50 Hz; 24VDC±10% |
Protection | IP65 |
Installation | Wall-mounted |
Size | 234/185/118mm, L/W/H |